Guidelines for Evaluation and
Selection of Continuous Hydrogen Sulfide Monitoring Equipment
INTRODUCTION
Section 10 is provided to make users of hydrogen sulfide monitoring equipment aware of some equipment limitations and certain desirable features of such equipment. There are a number of detection principles and analytical procedures available for monitoring the concentration of hydrogen sulfide in ambient air where the potential exists for exposure to levels that may be hazardous to health. These guidelines are intended to aid in the selection and application of continuous monitoring equipment for use in production operations involving hydrogen sulfide. The term "continuous hydrogen sulfide monitoring equipment" as used herein is defined as equipment capable of continuously measuring and displaying the concentration of hydrogen sulfide in ambient air. Section 10 is not applicable to personnel monitoring badges or length-of-stain or color-comparison type detector devices (refer to ANSI/ISEA 102).
GENERAL
All monitors, both portable and stationary, shall be designed on sound engineering and scientific principles and constructed of materials suitable for the application. Their design and construction should allow for ease of maintenance and repair. Instruments should be verified by a national recognized testing laboratory (NRTL) as meeting the minimum performance requirements of ISA-SI2.15, Part I. Equipment should be installed, operated, and maintained in accordance with ISA Recommended Practice 12.15, Part II.
It generally is recommended (and frequently required) that electrical controls for safety systems such as hydrogen sulfide monitoring equipment and other gas detector systems be installed normally-energized ("fail-safe"). This means that power is supplied continuously during normal operations to devices which provide alarm(s) and corrective action if concentrations corresponding to specific alarm set points are reached. Under these conditions, interruption of power due to either deliberate safety device actuation or loss of power will initiate corrective action. It is desirable to provide a test means that will allow the system to be tested (and calibrated) without shutting in producing or gas processing plant operations (or other corrective action), but it should be evident to operating personnel that the system is in the test (bypass) mode.
To better ensure proper application, it is recommended that an environmental and application checklist (similar to the example shown in Appendix 1, ISA Recommended Practice 12.15, Part II) be provided to prospective vendors by the
user.
CONSTRUCTION CHARACTERISTICS
The following construction and useability characteristics are desirable for hydrogen sulfide monitoring equipment.
Portability
Portable monitors, including all required parts and accessories, should weigh a maximum of ten (10) pounds and have a maximum volume of one (1) cubic foot.
Power Supply, Portable Monitoring
Equipment
Portable hydrogen sulfide monitoring equipment is defined as self-contained, battery-operated, carriable or transportable instruments capable of operating within specifications from integral batteries for a period of eight hours minimum, including a IS minute period of maximum load conditions (with alarms, lights, etc. activated), while exposed to clean air at a nominal temperature of 14°F (-10°C). Applications requiring in excess of eight hours of continuous operation or operation at lower temperatures should be specified by the end user.
Readout
Monitors should provide a direct readout of hydrogen sulfide concentration in parts per million (ppm) by volume.
Recorder Output
For certain applications, it may be desirable for monitors to provide an output signal (e.g., 4-20 ma) proportional to hydrogen sulfide concentration for use in connection with recorders or for other purposes.
Simplicity of Operation
Monitoring and detection equipment should be readily operable by personnel without scientific background or training in instrumentation.
Instruction Manuals
An instruction manual should be provided by the manufacturer with each instrument The instruction manual should contain complete operating instructions, including procedures for startup, warm-up time, zero checks, calibration, alarm setting and testing, preventive maintenance, performance checks, and trouble-shooting. Monitors with rechargeable power supplies should be furnished with in-
structions for charging, storing, and maintaining the power supply. Information also should be included regarding instrument recovery time after the exposure of sensor(s) to hydrogen sulfide. The manufacturer should provide response time data and a list of interfering, desensitizing, or contaminating substances or water vapor concentrations known to the instrument manufacturer which may adversely affect proper operation and performance of the instrument (refer to Par. 10.4.7). Instruction manuals should include wiring diagrams and estimates of the life expectancy of all consumables. The manual should include a complete parts list suitable for identification of all replaceable parts and sources for procurement of these parts.
Electrical Approval
Any portion of a stationary hydrogen sulfide monitoring instrument intended for installation or use in a hazardous (classified) location and all portable monitoring instruments shall be approved for use in such a hazardous (classified) location and marked accordingly. Refer to NFPA 70, Article 500-3.FPNNO.2.
Ruggedness
Portable monitoring units should be sufficiently rugged to withstand routine transporting, handling, and use in the field environment, as specified by the user. Refer to 1SA-S12.15 for details of a recommended "drop test" to evaluate portable unit ruggedness and a "vibration test" to evaluate ruggedness of fixed and portable monitors.
Calibration Equipment
All accessories required to calibrate the instruments should be made available by the manufacturer. The life expectancy and any special handling required of any hydrogen sulfide test concentrations should be provided by the supplier.
Zero and Span Adjustments
Zero and span adjustment controls should be readily accessible for field adjustment, and the monitor design should include provisions to apply zero and calibration gases to the sensor(s) in a non-laboratory environment. All accessories for calibration and zero should be made available with the monitor and should be useable under field conditions.
Alarm Systems
Fixed monitors shall have provisions for external alarms. Portable monitor units should contain integral audible, visual, or physical presentation (e.g., vibrator signal) alarms as specified by the user. Hydrogen sulfide alarms should be unique to the location.
Testing Alarm Circuitry
Provisions should be included for the testing of alarms and alarm outputs. The test procedure should be included in the equipment instruction manual.
Remote Sampling
A remote sampling accessory (such as a probe) may be desirable for use with a portable monitoring unit. CAUTION: Optional probe attachments for portable monitoring units, which allow the user to manually draw samples from remote locations, inherently prevent continuous monitoring of the immediate local environment. Users should consult the manufacturer's instruction manual to determine the proper number of bulb strokes required to draw samples when non-continuous monitoring attachments are used. The remote sampling attachment should be removed after use to restore the instrument to the normal continuous monitoring mode.
Equipment Trouble (Malfunction) Alarm
A trouble (malfunction) signal (indicator or output) should be provided for all monitors.
Detection Range Indication
The range(s) of detection should be conspicuously marked on the instrument.
PERFORMANCE GUIDELINES
The following recommended performance parameters are applicable to fixed and portable hydrogen sulfide monitoring equipment.
Accuracy
Instruments should meet requirements of the accuracy test specified in 1SA-S12.15, Part I. Users are cautioned that the class of instruments suitable for field use are not "laboratory-grade" instruments, and an equivalent degree of precision should not be expected.
Zero Drift
Instruments should meet the requirements of the "Long-term Stability Test" specified in ISA-S12.15, Part I. Excessive zero drift is undesirable and can require instrument calibration at unreasonably short intervals.
Warm-up Time
The minimum warm-up time when power is first applied should be stated in the equipment instruction manual. A monitor ready-status indicator is a desirable feature.
Response Time
ISA-S12.15, Part I specifies monitoring equipment minimum response time to input step changes. The toxicity of hydrogen sulfide requires that monitoring equipment have rapid response time to alert personnel of potentially dangerous concentrations. Hence, response time of monitoring equipment is an important parameter for consideration in evaluation and selection of such equipment.
Operating Humidity Range
Monitoring equipment should meet the "Humidity Variation Test" specified in ISA-SI2.15, Part I. Users should advise equipment manufacturers of the humidity ranges expected for specific equipment applications.
Operating Temperature Range
Monitoring equipment should be suitable for viable use in an ambient temperature range of 14°F to 122°F (-10°C to 50°C). Applications requiring equipment exposure to temperatures outside this range should be specified by the user.
Interferences
A list of interfering, desensitizing, and contaminating substances (e.g., carbon monoxide, sulfur dioxide, aromatic mercaptans, methanol, oxides of nitrogen, aldehydes, carbon sulfide, monoethanolamine, carbon dioxide, benzene, and methane) known by the manufacturer should be listed in the equipment instruction manual. Also, water vapor concentrations which may adversely affect proper equipment operation should be included in the instruction manual. CAUTION: Monitoring and detector equipment and sensors should be protected from exposure to liquid spray or wash down. Such exposure can affect equipment performance and reliability.
Functional Field Test
Functional field testing of monitoring equipment should be conducted under "as installed" or "as used" conditions. All instrument and system accessories normally used with this equipment should be installed and operated during functional field tests. Functional field tests may include, but are not limited to, exposing the sensor to a sample containing sufficient hydrogen sulfide to cause response of the system. Functional field testing does not necessarily include zero and span adjustments. The hydrogen sulfide concentration used in functional field testing of monitoring equipment should not exceed the maximum operating range of the equipment being tested.
Air Velocity
Monitoring equipment should meet the "Air Velocity Variation Test" specified in 1SA-S12.15, Part I. Accessories are
often available and may be desirable for use with detector equipment installed in areas of high air velocity.
Electromagnetic Interference (EMI)
Some monitoring equipment may be susceptible to electromagnetic interference (EMI), particularly radio frequency interference (RFI). Caution should be taken when using monitoring equipment in close proximity to a radio transmitter or other EMI generators. Refer to ISA-SI2.15, Part I, Sections 9.5 and 11.6.
Комментариев нет:
Отправить комментарий